Q. As a result of the experiment, amplification cannot be confirmed.
A. First, check that there is no problem with the template extraction process. Second, make sure that there are no missing contents (primer, probe, 2X PCR mix) when preparing the PCR mixture. Third, check the experimental conditions by reviewing the annealing temperature of the primer design and prime.
Q. As a result of the experiment, a signal was detected in the NC (negative control).
A. First, since non-specific reactions may occur due to environmental factors, please cleanly disinfect the equipment and experiment table you are using before proceeding. Second, since non-specific reactions may occur due to experimental factors (contamination by the template to be used), please make a mixture, dispense negative control first, and then proceed with the experiment on the target.
Q. Is ROX dye added to the product?
A. ROX dye is not added to this product, and it must be used after adding it if necessary. If there is a selection for passive reference dye during the experiment, please proceed with none.